Equinox Relieve, Analgesic Mixing Device

Description
Additional Info
Description

Designed for use in the pre-hospital (ambulance) and in-hospital use (ER, Labor and Delivery etc.), the Equinox® Relieve is intended to provide a 50/50% mixture of nitrous oxide and oxygen, on demand, to a conscious, spontaneously breathing, patient for the relief of pain due to trauma or other conditions where inhalation analgesia is clinically specified. The system has no operator or patient adjustable controls eliminating the risk of delivering a hypoxic mixture. It only requires the attachment to pressure-regulated medical gas oxygen and nitrous oxide cylinders and the patient circuit with mouthpiece and scavenging hose to be ready for use.

The Equinox® Relieve consists of a dual gas (Oxygen and Nitrous Oxide) control module mixing system that maintains the output gas mixture at a constant 50/50 mix. Each of the two inputs has specific built-in alarm systems that will generate both visual and audible alarms should either the nitrous oxide or oxygen input fall below 40 PSI. The device will then automatically shut off should either the nitrous oxide or oxygen input fall below 35 PSI.

The O-Two Equinox® Relieve is also equipped with a secondary “fail safe” circuit that will activate an alarm and shut off the device should an internal malfunction occur in the mixer or any internal hoses rupture or kink.

Equinox® Relieve N2O/O2 Analgesic Gas Mixing and Delivery System has only one control for turning ON or OFF the device. When it is turned ON, the output of the N2O/O2 gas mixture will only be activated by an inspiratory effort by the patient. The internal demand valve delivers high flow rates with low triggering pressures, reducing the amount of effort required by the patient to maintain the demanded flow. During use, the scavenging hose on the circuit assists in scavenging the patient’s exhaled air away from the patient’s environment.

The durable O-Two Equinox® Relieve N2O/O2 Analgesic Gas Mixing and Delivery System is a simple, safe and efficient way of delivering pain relief during trauma, childbirth or other painful procedures.

Additional Info
Minimum Flow rate

120 l/min

Exhalation Resistance

0 to 6 cmH2O @ 60 L/Min

Inhalation Resistance

0 to -6 cmH2O @ 60 L/Min

Oxygen Concentrations

42.5 – 57.5%y

Input pressure

50-70 PSI

Low Input Pressure Alarm

40 +/- 1 PSI

Device Cut Off Pressure

35 ± 4 PSI

Oxygen Enrichment Flow (O201EQ1000F only)

30 l/min

Demand Valve Triggering Pressure

1.5 to -2.5 cmH2O

Patient connector

15/22mm

Patient Circuit (Single Use)

Dual limb circuit with one-way valves and 19mm scavenging Port Connection on the exhalation limb (O201CV8028-CS)

Device Input Connections

Nitrous Oxide: CGA1040 DISS / Oxygen: CGA1240 DISS

Dimensions WxDxH

226 x 168 x 99mm

Weight

1.5kg

Ordering Info

O201EQ7000E – Equinox® II Administration System c/w Regulator, Bleed Pin, 7 Foot Supply Hose, Demand Valve, Universal Face Mask, Cylinder Wrench
O201EQ7000E – Equinox® II Administration System c/w Regulator (DISS female for cylinder config), Bleed Pin, Demand Valve, Universal Face Mask, Cylinder Wrench
O201EQ7001E – Equinox® II Demand Valve c/w 7 Foot Supply Hose, Universal Resuscitation Face Mask
O201EQ7001E-BS – Equinox® II Demand Valve c/w 7 Foot Supply Hose c/w BS5682 Probe, Universal Resuscitation Face Mask

Consumables:
O201CV8028-CS – Single-Use 6 Foot Circuit with Scavenger Hose and Adapter with Mouthpiece, Case/10