Newly designed, the modules are available in the following: Semi-Flush or Surface-Mounted.
- HTM 02-01, BS EN 18082, and C11 fully compliant ensuring individual gas separation in case of emergency.
- Designed to fit ceiling heights of 2.4mtr, 2.7mtr or 3mtr as standard, other lengths available.
- Gas venting as per BS EN 11197 to avoid inadvertent gas build-up in cavity wall installation.
- Upstream and Downstream NIST connectors with 100% serviceable non-return valves fitted on all gases.
- Gas types available O2, N2O,O2/N2O, MA4, SA7, VAC, CO2, N2
- Factory tested and certified Pre-Piped EN 13348 Copper Tube
- Cover types available- Epoxy-coated steel or Polymer laminated anti-Microbial cover.
- Extremely hard-wearing, impact-proof Polymer Laminate material, available in a selection of colours.
- Factory Fitted CPX Analogue or Digital Area Alarm (Also available to be manufactured as remote alarm set-up)
- Onboard pressure sensors shall be fitted local to each gas line behind a locked door. Integrated 12V Digital Alarm factory set for the Gas types and Pressure Settings. Battery backup in case of power failure shall continue to operate for up to 24 hours.
- Option Flow Measurement can be connected to each individual gas type to provide flow data 24 hours each day.
- Delivered assembled to site for easy one-man installation.